INSTRUMENTATION LABORATORY CO. FDA Inspection Record
INSTRUMENTATION LABORATORY CO. is an FDA-registered facility (FEI 1217183) located in Bedford, MA, US. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jul 12, 2013: Voluntary Action Indicated (VAI)
- Inspections on record
- 2 (2 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 25, latest Sep 3, 2026
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 12, 2013 | Voluntary Action Indicated (VAI) | No citations listed |
| Aug 24, 2011 | Voluntary Action Indicated (VAI) | Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) Supplier notification of changes (21 CFR 820.50(b)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) Evaluation of suppliers, contractors, etc., requirements (21 CFR 820.50(a)) and 1 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 3, 2026 | Class I | Z-0001-2027 | HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic and 1 more product Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples. |
| May 21, 2026 | Class II | Z-2899-2026 | GEM Premier 7000 with iQM3
Part Numbers:
00077407508
00077407510
00077415004
000… and 1 more product The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established… |
| May 8, 2026 | Class II | Z-2821-2026 | GEM Premier 5000 with IQM2
Model/Catalog Number: 00024019255 and
GEM Premier 7000 with… and 5 more products Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an… |
| Jan 20, 2026 | Class II | Z-1524-2026 | GEM Premier 5000 PAK, Part No. 00055360004. and 19 more products Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of… |
| Dec 11, 2025 | Class II | Z-1095-2026 | The product is used for the evaluation of the intrinsic coagulation pathway, APTT… Potential for microbial contamination. |
| Nov 18, 2025 | Class III | Z-0938-2026 | HemosIL SynthAFax. Partial Thromboplastin Time Tests. Recalled lots were manufactured with double the amount of preservative concentration. |
| Feb 3, 2025 | Class II | Z-2323-2025 | HemosIL LMW Heparin Controls;
Part Number: 0020300200; Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of… |
| Dec 18, 2024 | Class II | Z-1020-2025 | HemosIL Heparin Calibrators; Part Number: 0020300600; in vitro diagnostic Control indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package… |
| Sep 29, 2023 | Class II | Z-0357-2024 | GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004… and 2 more products Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the… |
| Jul 13, 2023 | Class II | Z-2578-2023 | ACL TOP Family 50 Series Models as follows: In vitro diagnostic clinical use in the… Sample misidentification could occur under specific conditions and patient management altered based on an incorrectly assigned result |
FDA record score
56 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
26 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-0001-2027, this FEI as the recalling firm | Sep 3, 2026 | 15 x 1 | 15 |
| Class II recall Z-2899-2026, this FEI as the recalling firm | May 21, 2026 | 4 x 1 | 4 |
| Class II recall Z-2821-2026, this FEI as the recalling firm | May 8, 2026 | 4 x 1 | 4 |
| Class II recall Z-1524-2026, this FEI as the recalling firm | Jan 20, 2026 | 4 x 1 | 4 |
| Class II recall Z-1095-2026, this FEI as the recalling firm | Dec 11, 2025 | 4 x 1 | 4 |
| Class II recall Z-2323-2025, this FEI as the recalling firm | Feb 3, 2025 | 4 x 1 | 4 |
| Class II recall Z-1020-2025, this FEI as the recalling firm | Dec 18, 2024 | 4 x 1 | 4 |
| Class II recall Z-0357-2024, this FEI as the recalling firm | Sep 29, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-2578-2023, this FEI as the recalling firm | Jul 13, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1304-2023, this FEI as the recalling firm | Feb 3, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1052-2022, this FEI as the recalling firm | Apr 8, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1014-2022, this FEI as the recalling firm | Feb 22, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0276-2022, this FEI as the recalling firm | Oct 13, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0149-2022, this FEI as the recalling firm | Aug 25, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2485-2019, this FEI as the recalling firm | Jul 23, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0077-2019, this FEI as the recalling firm | Aug 1, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0895-2018, this FEI as the recalling firm | Aug 2, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0736-2016, this FEI as the recalling firm | Dec 18, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0832-2015, this FEI as the recalling firm | Nov 21, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2060-2014, this FEI as the recalling firm | Jun 3, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0514-2014, this FEI as the recalling firm | Nov 15, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2244-2013, this FEI as the recalling firm | Jul 19, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 12, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1784-2013, this FEI as the recalling firm | Apr 9, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2387-2012, this FEI as the recalling firm | Aug 3, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 24, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 14, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 15, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing INSTRUMENTATION LABORATORY CO. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check INSTRUMENTATION LABORATORY CO. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Instrumentation Laboratory
- Location
- Bedford, MA (US)
- Source
- device