FDA Inspection 742203 — Instrumentation Laboratory

Instrumentation Laboratory — FEI 1217183

The FDA completed an inspection of Instrumentation Laboratory on Aug 24, 2011 in Bedford, MA. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 5 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
Aug 24, 2011
Fiscal Year
2011
Product Type
Devices
Project Area
Compliance: Devices
Location
Bedford, MA (United States)

Citations

IDCFRDescription
312121 CFR 820.80(b)Lack of or inadequate receiving acceptance procedures
320721 CFR 820.50(b)Supplier notification of changes
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
48621 CFR 820.50(a)Evaluation of suppliers, contractors, etc., requirements
54621 CFR 820.75(a)Lack of or inadequate process validation

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.