APPLIED MEDICAL RESOURCES CORP. FDA Inspection Record
APPLIED MEDICAL RESOURCES CORP. is an FDA-registered facility (FEI 2027111) located in RSM, CA, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Sep 2, 2021: No Action Indicated (NAI)
- Inspections on record
- 5 (2 OAI, 1 VAI, 2 NAI)
- Last OAI inspection
- Oct 14, 2015
- Warning letters
- 1 listed by FDA for this FEI, latest Apr 10, 2015
- Recalls as the recalling firm
- 15, latest Dec 11, 2025
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 2, 2021 | No Action Indicated (NAI) | No citations listed |
| Nov 20, 2018 | No Action Indicated (NAI) | No citations listed |
| Mar 31, 2017 | Voluntary Action Indicated (VAI) | No citations listed |
| Oct 14, 2015 | Official Action Indicated (OAI) | Design verification - output does not meet input requirement (21 CFR 820.30(f)) Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) and 4 more |
| Dec 16, 2014 | Official Action Indicated (OAI) | Design input - Lack of or inadequate procedures (21 CFR 820.30(c)) Design verification - output does not meet input requirement (21 CFR 820.30(f)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) and 4 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Dec 11, 2025 | Class II | Z-1010-2026 | Kii Low Profile, Bladed Dual Pack , Qty. 1 5x55mm
REF: CTB23 There is the potential for extended length of the blade obturator after assembly into the cannula and seal housing. |
| Jan 30, 2024 | Class II | Z-1388-2024 | Epix Universal Clip Applier, REF: CA500, STERILEEO and 1 more product Their is the potential that Universal Clip applier may not load clip after the trigger is actuated. |
| Mar 17, 2020 | Class II | Z-2158-2020 | Epix latis GRASPER, 5 mm x 35 cm, REF C4130, Qty: 10, Sterile R, CE, Rx ONLY, UDI… and 3 more products There may be a slightly protruding rivet on the distal end of the laparoscopic grasper which has the potential for tissue to catch on it… |
| Jan 31, 2020 | Class II | Z-2034-2020 | STEALTH SPRING CLIP 6mm latis (1/2 Force). REF/UDI::A1601/(01)00607915110567 - Product… and 3 more products Failure of cardiovascular clips to provide sufficient occlusion of the vessel. |
| Oct 24, 2019 | Class I | Z-1846-2020 | Python Embolectomy Catheters, Sterile EO, CE, Ref numbers/UDI:
A4E01/(01)00607915113407… and 2 more products The firm has received reports of tip separation during usage of the catheters, which may expose patients to unintended dislodgement of the… |
| Jan 28, 2019 | Class II | Z-1959-2019 | Kii Fios First Entry, 5 X 75 mm, RX only, REF CFF05 Qty: 6, Sterile R, Lot: 1330981, Exp… The product may not have met sterility requirements . Use of a non-sterile device on a patient may expose the patient to infectious agents. |
| Dec 12, 2018 | Class II | Z-0509-2020 | Epix Electrosurgical Probes with Smoke Evacuation, Angled L-Hook Tip, 5mm x 42 cm
REF… and 1 more product Unintentional shifting of the tip insulation of the firm's electrosurgical probes. |
| May 4, 2017 | Class II | Z-2533-2017 | Kii Shielded Bladed trocars (CTB14), Kii Low profile trocars (CTB23), and
Kii 5x55mm… Due to possibility of insufficient sterilization. |
| Apr 1, 2016 | Class II | Z-1621-2016 | Direct Drive Clip Applier CA090, indicated for ligation of tubular structures or vessels… Increased customer complaints indicating inconsistent clip application; may lead to unoccluded vessels. |
| Dec 22, 2015 | Class II | Z-1329-2016 | CA500 Epix Universal Clip Applier
Product Usage:
The Epix disposable laparoscopic… and 3 more products Applied Medical Resources Corporation is recalling CA500 Epix Universal Clip Applier, kit models STD GK258 AME Standard Kit 44, STD KIT… |
FDA record score
17 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
19 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1010-2026, this FEI as the recalling firm | Dec 11, 2025 | 4 x 1 | 4 |
| Class I recall Z-1846-2020, this FEI as the recalling firm | Oct 24, 2019 | 15 x 0.25 | 3.8 |
| Class II recall Z-1388-2024, this FEI as the recalling firm | Jan 30, 2024 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Mar 31, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-2158-2020, this FEI as the recalling firm | Mar 17, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2034-2020, this FEI as the recalling firm | Jan 31, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1959-2019, this FEI as the recalling firm | Jan 28, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0509-2020, this FEI as the recalling firm | Dec 12, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2533-2017, this FEI as the recalling firm | May 4, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1621-2016, this FEI as the recalling firm | Apr 1, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1329-2016, this FEI as the recalling firm | Dec 22, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Oct 14, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2769-2015, this FEI as the recalling firm | Jul 31, 2015 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Apr 10, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Dec 16, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0183-2014, this FEI as the recalling firm | Oct 21, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1521-2013, this FEI as the recalling firm | May 1, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0644-2013, this FEI as the recalling firm | Nov 5, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2446-2012, this FEI as the recalling firm | Aug 9, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 24, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 24, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 24, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |