FDA Inspection 908051 — Applied Medical Resources Corp

Applied Medical Resources Corp — FEI 2027111

The FDA completed an inspection of Applied Medical Resources Corp on Dec 16, 2014 in Rancho Santa Margarita, CA. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 8 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Official Action Indicated (OAI)
End Date
Dec 16, 2014
Fiscal Year
2015
Product Type
Devices
Project Area
Compliance: Devices
Location
Rancho Santa Margarita, CA (United States)

Citations

IDCFRDescription
1471421 CFR 820.30(c)Design input - Lack of or inadequate procedures
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
313021 CFR 820.100(a)Lack of or inadequate procedures
317221 CFR 820.198(c)Investigation of device failures
319121 CFR 820.30(g)Design validation - production units
367621 CFR 820.30(f)Design verification - documentation
419221 CFR 806.10(a)(2)Report of violation of the Act (see 803.52(e)(9))
73221 CFR 803.50(a)(2)Individual Report of Malfunction