Wassenburg Medical, Inc. FDA Inspection Record

FEI 2523209 · Inspections, Form 483s, warning letters and import refusals

Wassenburg Medical, Inc. is an FDA-registered facility (FEI 2523209) located in Ivyland, PA, US. RegDataLab tracks 7 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Oct 19, 2018: No Action Indicated (NAI)
Inspections on record
7 (3 OAI, 2 VAI, 2 NAI)
Last OAI inspection
Apr 24, 2015
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
2, latest May 6, 2016

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Oct 19, 2018No Action Indicated (NAI)No citations listed
Apr 28, 2017No Action Indicated (NAI)No citations listed
Apr 24, 2015Official Action Indicated (OAI)Lack of or inadequate process validation (21 CFR 820.75(a))
Nov 19, 2014Voluntary Action Indicated (VAI)Design verification acceptance criteria (21 CFR 820.30(f))
Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h))
Lack of or inadequate In-process acceptance procedures (21 CFR 820.80(c))
Acceptance records (21 CFR 820.184(d))
and 4 more
Nov 8, 2012Official Action Indicated (OAI)No citations listed
Jan 26, 2012Official Action Indicated (OAI)DHR - not or inadequately maintained (21 CFR 820.184)
Design verification - output does not meet input requirement (21 CFR 820.30(f))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Complaints (21 CFR 820.198(a))
and 5 more
May 26, 2010Voluntary Action Indicated (VAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
May 6, 2016Class IIZ-2130-2016The System 83 Plus, Endoscope Washer/Disinfector Designed for the simultaneous…
To warn customers that Custom Ultrasonics, Inc. System 83 Plus AERs should not be used for cleaning and/or high-level disinfection of…
Mar 12, 2013Class IIZ-0426-2014Custom Ultrasonics System 83 Plus 2 and Plus 9 Endoscope Washer/Disinfector.
Certain System 83 Plus 2 and System 83 Plus 9 units running the Windows 7 operating system were not fully validated prior to distribution.

FDA record score

0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 28, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2130-2016, this FEI as the recalling firmMay 6, 201610 years or older, so it adds no points0
Inspection classified OAIApr 24, 201510 years or older, so it adds no points0
Inspection classified VAINov 19, 201410 years or older, so it adds no points0
Class II recall Z-0426-2014, this FEI as the recalling firmMar 12, 201310 years or older, so it adds no points0
Inspection classified OAINov 8, 201210 years or older, so it adds no points0
Inspection classified OAIJan 26, 201210 years or older, so it adds no points0
Inspection classified VAIMay 26, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 28, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 28, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 28, 2026Newest record Nov 19, 2026Current
Warning letter textNov 28, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 28, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Wassenburg Medical, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Wassenburg Medical, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Wassenburg Medical, Inc. → Check this supplier

Facility details

Company
Wassenburg Medical, Inc.
Location
Ivyland, PA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.