Wassenburg Medical, Inc. FDA Inspection Record
Wassenburg Medical, Inc. is an FDA-registered facility (FEI 2523209) located in Ivyland, PA, US. RegDataLab tracks 7 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Oct 19, 2018: No Action Indicated (NAI)
- Inspections on record
- 7 (3 OAI, 2 VAI, 2 NAI)
- Last OAI inspection
- Apr 24, 2015
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest May 6, 2016
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 19, 2018 | No Action Indicated (NAI) | No citations listed |
| Apr 28, 2017 | No Action Indicated (NAI) | No citations listed |
| Apr 24, 2015 | Official Action Indicated (OAI) | Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Nov 19, 2014 | Voluntary Action Indicated (VAI) | Design verification acceptance criteria (21 CFR 820.30(f)) Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h)) Lack of or inadequate In-process acceptance procedures (21 CFR 820.80(c)) Acceptance records (21 CFR 820.184(d)) and 4 more |
| Nov 8, 2012 | Official Action Indicated (OAI) | No citations listed |
| Jan 26, 2012 | Official Action Indicated (OAI) | DHR - not or inadequately maintained (21 CFR 820.184) Design verification - output does not meet input requirement (21 CFR 820.30(f)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Complaints (21 CFR 820.198(a)) and 5 more |
| May 26, 2010 | Voluntary Action Indicated (VAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| May 6, 2016 | Class II | Z-2130-2016 | The System 83 Plus, Endoscope Washer/Disinfector
Designed for the simultaneous… To warn customers that Custom Ultrasonics, Inc. System 83 Plus AERs should not be used for cleaning and/or high-level disinfection of… |
| Mar 12, 2013 | Class II | Z-0426-2014 | Custom Ultrasonics System 83 Plus 2 and Plus 9 Endoscope Washer/Disinfector. Certain System 83 Plus 2 and System 83 Plus 9 units running the Windows 7 operating system were not fully validated prior to distribution. |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2130-2016, this FEI as the recalling firm | May 6, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Apr 24, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Nov 19, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0426-2014, this FEI as the recalling firm | Mar 12, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Nov 8, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Jan 26, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | May 26, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 28, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 28, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 28, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 28, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 28, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Wassenburg Medical, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).