FDA Inspection 765157 — Wassenburg Medical, Inc.

Wassenburg Medical, Inc. — FEI 2523209

The FDA completed an inspection of Wassenburg Medical, Inc. on Jan 26, 2012 in Ivyland, PA. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 9 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Official Action Indicated (OAI)
End Date
Jan 26, 2012
Fiscal Year
2012
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Ivyland, PA (United States)

Citations

IDCFRDescription
1471221 CFR 820.184DHR - not or inadequately maintained
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
316821 CFR 820.198(a)Complaints
319221 CFR 820.30(g)Design validation - user needs and intended uses
337221 CFR 820.198(d)Records of MDR Investigation
368321 CFR 820.70(g)Equipment Installation, Placement, Specified Requirements
405721 CFR 820.20(a)Management ensuring quality policy is understood
407021 CFR 820.30(g)Design validation - documentation

Need the full FDA record for Wassenburg Medical, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Wassenburg Medical, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Wassenburg Medical, Inc. →

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.