Xoran Technologies, LLC FDA Inspection Record
Xoran Technologies, LLC is an FDA-registered facility (FEI 3004198450) located in Ann Arbor, MI, US. RegDataLab tracks 7 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jul 11, 2024: Official Action Indicated (OAI)
- Inspections on record
- 7 (4 OAI, 3 VAI)
- Last OAI inspection
- Jul 11, 2024
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Dec 9, 2024. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 6, latest Sep 8, 2014
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 11, 2024 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design validation - software validation not performed (21 CFR 820.30(g)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| Dec 20, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Justification for not reporting (21 CFR 806.20(b)(4)) |
| Oct 6, 2014 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 5 more |
| May 14, 2012 | Official Action Indicated (OAI) | Lack/Inad procedure-Monitoring/Control of Validated Proces (21 CFR 820.75(b)) Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) Documentation (21 CFR 820.80(e)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 15 more |
| Apr 29, 2011 | Official Action Indicated (OAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) Lack of or inadequate procedures (21 CFR 820.100(a)) Rationale documented for no investigation (21 CFR 820.198(b)) and 27 more |
| Oct 14, 2010 | Official Action Indicated (OAI) | Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Lack of or inadequate procedures (21 CFR 820.100(a)) Review and evaluation for investigation (21 CFR 820.198(b)) Management ensuring quality policy is understood (21 CFR 820.20(a)) and 4 more |
| Mar 4, 2009 | Voluntary Action Indicated (VAI) | Establish the design and development plan (21 CFR 820.30(b)) Design output - review and approval (21 CFR 820.30(d)) Design review - documentation (21 CFR 820.30(e)) Design verification procedures (21 CFR 820.30(f)) and 15 more |
Warning letters
- Dec 9, 2024: CGMP/QSR/Medical Devices/Adulterated/Misbranded. Read the letter on FDA.gov.
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 8, 2014 | Class II | Z-1210-2015 | Xoran MiniCAT X-ray imaging device Out of tolerance kVp, Unexposed (black) frames, and Out of tolerance Dose. |
| Jul 27, 2010 | Class II | Z-0036-2014 | Xoran Mini CT Scanner A description of the defect in the product or the manner in which the product fails to comply with an applicable Federal standard; The… |
| Jul 27, 2010 | Class II | Z-0037-2014 | Xoran MiniCAT CT Scanner A description of the defect in the product or the manner in which the product fails to comply wvith an applicable Federal standard; The… |
| Jul 27, 2010 | Class II | Z-0038-2014 | Xoran Minicat00202 and 1 more product The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints… |
| Jul 27, 2010 | Class II | Z-0040-2014 | Xoran MiniCAT00202 The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints… |
| Dec 14, 2009 | Class II | Z-0041-2014 | Xoran MiniCAT CT Scanner ID ENT00001 Failure of the gantry to rotate. The complaint alleged that the scanner would emit radiation but would not rotate. |
FDA record score
50 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
14 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Warning letter, no close-out letter listed | Dec 9, 2024 | 30 x 1 | 30 |
| Inspection classified OAI | Jul 11, 2024 | 30 x 0.6 | 18 |
| Inspection classified VAI | Dec 20, 2017 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Oct 6, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1210-2015, this FEI as the recalling firm | Sep 8, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | May 14, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Apr 29, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Oct 14, 2010 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0036-2014, this FEI as the recalling firm | Jul 27, 2010 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0037-2014, this FEI as the recalling firm | Jul 27, 2010 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0038-2014, this FEI as the recalling firm | Jul 27, 2010 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0040-2014, this FEI as the recalling firm | Jul 27, 2010 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0041-2014, this FEI as the recalling firm | Dec 14, 2009 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 4, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Xoran Technologies, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.