558909

Xoran Technologies, LLC — FEI 3004198450

Inspection Details

Classification
Voluntary Action Indicated (VAI) (VAI)
End Date
March 4, 2009
Fiscal Year
2009
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Ann Arbor, MI (United States)

Citations

IDCFRDescription
242921 CFR 820.30(b)Establish the design and development plan
260221 CFR 820.30(d)Design output - review and approval
263021 CFR 820.30(e)Design review - documentation
264821 CFR 820.30(f)Design verification procedures
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
310221 CFR 820.30(h)Design transfer - Lack of or inadequate procedures
315521 CFR 820.181(a)DMR device specifications
315921 CFR 820.184DHR content
320121 CFR 820.40(a)Not approved or obsolete document retrieval
320821 CFR 820.50(b)Approval of purchasing data {see also 820.40(a)}
328421 CFR 820.90(a)Nonconforming product evaluation/investigation
331021 CFR 820.120(b)DHR documentation of label release {see also 820.184}
333121 CFR 820.181DMR - not or inadequately maintained
333221 CFR 820.184(e)ID label, labeling
341921 CFR 820.70(a)(4)Approval of process
367321 CFR 820.30(c)Design input approval documentation
367721 CFR 820.30(g)Design validation - software validation not performed
405721 CFR 820.20(a)Management ensuring quality policy is understood
407121 CFR 820.30(i)Design changes - validation vs. verification