1010653

Forward Science Technologies LLC — FEI 3010143501

Inspection Details

Classification
Voluntary Action Indicated (VAI) (VAI)
End Date
November 30, 2016
Fiscal Year
2017
Product Type
Devices
Project Area
Compliance: Devices
Location
Houston, TX (United States)

Citations

IDCFRDescription
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
328521 CFR 820.90(b)(2)Product rework procedures, Lack of or inadequate procedures
63221 CFR 803.17(a)(2)Lack of System for Determining MDR Events