FDA Inspection 1029762 — Delta Med SpA

Delta Med SpA — FEI 3006846316

The FDA completed an inspection of Delta Med SpA on Oct 26, 2017 in Viadana. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 6 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
Oct 26, 2017
Fiscal Year
2018
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Viadana (Italy)

Citations

IDCFRDescription
260421 CFR 820.30(e)Design review - Lack of or inadequate procedures
265021 CFR 820.30(f)Design verification - Lack of or inadequate procedures
311721 CFR 820.70(i)Software validation for automated processes
326321 CFR 820.250(b)Sampling plans
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation

Need the full FDA record for Delta Med SpA? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key