FDA Inspection 1031661 — Bioptron AG

Bioptron AG — FEI 3003775404

The FDA completed an inspection of Bioptron AG on Nov 2, 2017 in Wollerau. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 12 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
Nov 2, 2017
Fiscal Year
2018
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Wollerau (Switzerland)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
255721 CFR 820.30(c)Design input - documentation
260421 CFR 820.30(e)Design review - Lack of or inadequate procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
340921 CFR 820.200(d)Service reports
342721 CFR 820.50(a)(2)Supplier oversight
367621 CFR 820.30(f)Design verification - documentation
407021 CFR 820.30(g)Design validation - documentation
569421 CFR 1002.10(d)Description of function, op. characteristics, uses
63121 CFR 803.17(a)(1)Lack of System for Event Evaluations

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.