FDA Inspection 1032463 — Biomonde US, LLC
The FDA completed an inspection of Biomonde US, LLC on Nov 17, 2017 in Gainesville, FL. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 7 citations were issued during this inspection. Product area: Devices.
Inspection Details
- Classification
- Voluntary Action Indicated (VAI)
- End Date
- Nov 17, 2017
- Fiscal Year
- 2018
- Product Type
- Devices
- Project Area
- Project Evaluation: Devices
- Location
- Gainesville, FL (United States)
Citations
| ID | CFR | Description |
|---|---|---|
| 3104 | 21 CFR 820.30(j) | Design history file |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3678 | 21 CFR 820.30(g) | Design Validation - Risk analysis not performed/inadequate |
| 479 | 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures |
| 541 | 21 CFR 820.70(c) | Environmental control Lack of or inadequate procedures |
| 546 | 21 CFR 820.75(a) | Lack of or inadequate process validation |
| 630 | 21 CFR 803.17 | Lack of Written MDR Procedures |