FDA Inspection 1039256 — CL Medical

CL Medical — FEI 3005347805

The FDA completed an inspection of CL Medical on Feb 1, 2018 in Ste Foy Les Lyon. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 5 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
Feb 1, 2018
Fiscal Year
2018
Product Type
Devices
Project Area
Compliance: Devices
Location
Ste Foy Les Lyon (France)

Citations

IDCFRDescription
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
326421 CFR 820.250(b)Sampling methods - Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
73121 CFR 803.50(a)(1)Report of Death or Serious Injury

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.