FDA Inspection 1062806 — PCCI, Inc.
The FDA completed an inspection of PCCI, Inc. on August 16, 2018 in Alexandria, VA. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions were found but do not meet the threshold for regulatory action). 13 citations were issued during this inspection. Product area: Devices.
Inspection Details
- Classification
- Voluntary Action Indicated (VAI) (VAI)
- End Date
- August 16, 2018
- Fiscal Year
- 2018
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Alexandria, VA (United States)
Citations
| ID | CFR | Description |
|---|---|---|
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 2350 | 21 CFR 820.25(b) | Training - Lack of or inadequate procedures |
| 3103 | 21 CFR 820.30(i) | Design changes - Lack of or Inadequate Procedures |
| 3120 | 21 CFR 820.80(a) | Lack of or inadequate procedures - Acceptance activities |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3132 | 21 CFR 820.120 | Lack of or inadequate procedures for labeling |
| 3233 | 21 CFR 820.72(a) | Calibration, Inspection, etc. Procedures Lack of or Inadequ |
| 3282 | 21 CFR 820.90(a) | Nonconforming product, Lack of or inadequate procedures |
| 3669 | 21 CFR 820.20(c) | Management review - defined interval, sufficient frequency |
| 3678 | 21 CFR 820.30(g) | Design Validation - Risk analysis not performed/inadequate |
| 3680 | 21 CFR 820.70(a) | Process control procedures, Lack of or inadequate procedures |
| 4059 | 21 CFR 820.22 | Quality Audits - defined intervals |
| 479 | 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures |