1062806

PCCI, Inc. — FEI 3004035771

Inspection Details

Classification
Voluntary Action Indicated (VAI) (VAI)
End Date
August 16, 2018
Fiscal Year
2018
Product Type
Devices
Project Area
Compliance: Devices
Location
Alexandria, VA (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
235021 CFR 820.25(b)Training - Lack of or inadequate procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
313021 CFR 820.100(a)Lack of or inadequate procedures
313221 CFR 820.120Lack of or inadequate procedures for labeling
323321 CFR 820.72(a)Calibration, Inspection, etc. Procedures Lack of or Inadequ
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
366921 CFR 820.20(c)Management review - defined interval, sufficient frequency
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
368021 CFR 820.70(a)Process control procedures, Lack of or inadequate procedures
405921 CFR 820.22Quality Audits - defined intervals