FDA Inspection 1092639 — Wuxi Contact Electronics Co., Ltd.
The FDA completed an inspection of Wuxi Contact Electronics Co., Ltd. on May 16, 2019 in Wuxi. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 8 citations were issued during this inspection. Product area: Devices.
Inspection Details
- Classification
- Official Action Indicated (OAI) (OAI)
- End Date
- May 16, 2019
- Fiscal Year
- 2019
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Wuxi (China)
Citations
| ID | CFR | Description |
|---|---|---|
| 14722 | 21 CFR 820.40 | Procedures not adequately established or maintained |
| 2327 | 21 CFR 820.22 | Quality audits - Lack of or inadequate procedures |
| 2350 | 21 CFR 820.25(b) | Training - Lack of or inadequate procedures |
| 3120 | 21 CFR 820.80(a) | Lack of or inadequate procedures - Acceptance activities |
| 3233 | 21 CFR 820.72(a) | Calibration, Inspection, etc. Procedures Lack of or Inadequ |
| 3263 | 21 CFR 820.250(b) | Sampling plans |
| 3680 | 21 CFR 820.70(a) | Process control procedures, Lack of or inadequate procedures |
| 546 | 21 CFR 820.75(a) | Lack of or inadequate process validation |