1110467

Biomedix Wai — FEI 1000115518

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
October 31, 2019
Fiscal Year
2020
Product Type
Devices
Project Area
Compliance: Devices
Location
Chanhassen, MN (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
311921 CFR 820.75(b)Lack/Inad procedure-Monitoring/Control of Validated Proces
312321 CFR 820.80(c)Lack of or inadequate In-process acceptance procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
322621 CFR 820.70(g)(1)Maintenance schedule, Lack of or inadequate schedule
326321 CFR 820.250(b)Sampling plans
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
54121 CFR 820.70(c)Environmental control Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation