FDA Inspection 1155101 — Tandem Diabetes Care, Inc.

Tandem Diabetes Care, Inc. — FEI 3007981285

The FDA completed an inspection of Tandem Diabetes Care, Inc. on October 15, 2021 in San Diego, CA. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions were found but do not meet the threshold for regulatory action). 4 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI) (VAI)
End Date
October 15, 2021
Fiscal Year
2022
Product Type
Devices
Project Area
Compliance: Devices
Location
San Diego, CA (United States)

Citations

IDCFRDescription
313021 CFR 820.100(a)Lack of or inadequate procedures
317221 CFR 820.198(c)Investigation of device failures
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
419121 CFR 806.10(a)(1)Report of risk to health