1216506
O3UV, LLC — FEI 3027918571
Inspection Details
- Classification
- Official Action Indicated (OAI) (OAI)
- End Date
- August 31, 2023
- Fiscal Year
- 2023
- Product Type
- Devices
- Project Area
- Postmarket Assurance: Devices
- Location
- Grand Ledge, MI (United States)
Citations
| ID | CFR | Description |
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 2327 | 21 CFR 820.22 | Quality audits - Lack of or inadequate procedures |
| 2371 | 21 CFR 820.30(a) | Design control - no procedures |
| 3120 | 21 CFR 820.80(a) | Lack of or inadequate procedures - Acceptance activities |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3160 | 21 CFR 820.184 | Lack of or inadequate DHR procedures |
| 3666 | 21 CFR 820.20(c) | Management review - Lack of or inadequate procedures |
| 3680 | 21 CFR 820.70(a) | Process control procedures, Lack of or inadequate procedures |
| 479 | 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures |