FDA Inspection 1232238 — Precision Diabetes Inc.

Precision Diabetes Inc. — FEI 3016698405

The FDA completed an inspection of Precision Diabetes Inc. on Mar 8, 2024 in Raleigh, NC. The inspection was classified NAI (No Action Indicated — no objectionable conditions or practices were found during the inspection). 7 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
No Action Indicated (NAI)
End Date
Mar 8, 2024
Fiscal Year
2024
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Raleigh, NC (United States)

Citations

IDCFRDescription
237121 CFR 820.30(a)Design control - no procedures
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
313021 CFR 820.100(a)Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
366621 CFR 820.20(c)Management review - Lack of or inadequate procedures
405921 CFR 820.22Quality Audits - defined intervals
48621 CFR 820.50(a)Evaluation of suppliers, contractors, etc., requirements

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.