1259482

Exactech, Inc. — FEI 1038671

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
August 27, 2025
Fiscal Year
2025
Product Type
Devices
Project Area
Compliance: Devices
Location
Gainesville, FL (United States)

Citations

IDCFRDescription
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
255721 CFR 820.30(c)Design input - documentation
313021 CFR 820.100(a)Lack of or inadequate procedures
333121 CFR 820.181DMR - not or inadequately maintained
343321 CFR 820.75(c)Process changes - review, evaluation and revalidation
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
53921 CFR 820.70(b)Production and Process Change Procedures, lack of or Inad.
54621 CFR 820.75(a)Lack of or inadequate process validation