FDA Inspection 1261809 — Sewoon Medical Co., Ltd.
The FDA completed an inspection of Sewoon Medical Co., Ltd. on Feb 20, 2025 in Cheonan. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 5 citations were issued during this inspection. Product area: Devices.
Inspection Details
- Classification
- Official Action Indicated (OAI)
- End Date
- Feb 20, 2025
- Fiscal Year
- 2025
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Cheonan (Korea (the Republic of))
Citations
| ID | CFR | Description |
|---|---|---|
| 14715 | 21 CFR 820.30(d) | Design output - Lack of or inadequate procedures |
| 3119 | 21 CFR 820.75(b) | Lack/Inad procedure-Monitoring/Control of Validated Proces |
| 3678 | 21 CFR 820.30(g) | Design Validation - Risk analysis not performed/inadequate |
| 3680 | 21 CFR 820.70(a) | Process control procedures, Lack of or inadequate procedures |
| 546 | 21 CFR 820.75(a) | Lack of or inadequate process validation |