ECLOSION — FEI 3030198312

The FDA completed an inspection of ECLOSION on Aug 13, 2026 in Fairlawn, OH. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 10 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
Aug 13, 2026
Fiscal Year
2026
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Fairlawn, OH (United States)

Citations

IDCFRDescription
2280021 CFR 830.300(a)Devices subject to device identification GUDID data submission requirements.
23038ISO 13485:2016 Clause 4.2.4Lack of document control
23085ISO 13485:2016 Clause 7.1Processes needed for product realization not planned and developed
23136ISO 13485:2016 Clause 7.3.10Design and development files not maintained
23138ISO 13485:2016 Clause 7.4.1Procedures for purchasing process not documented
23150ISO 13485:2016 Clause 7.4.3Purchased product verification activities not established or implemented
23244ISO 13485:2016 Clause 8.3.3Records not maintained for nonconforming product actions detected after delivery
23258ISO 13485:2016 Clause 8.5.2Corrective action procedures not documented
23261ISO 13485:2016 Clause 8.5.3Preventive action procedures not documented
63021 CFR 803.17Lack of Written MDR Procedures