FDA Inspection 1335100 — Walz Elektronik Gmbh
The FDA completed an inspection of Walz Elektronik Gmbh on Sep 10, 2026 in Rohrdorf. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 3 citations were issued during this inspection. Product area: Devices.
Inspection details
- Classification
- Voluntary Action Indicated (VAI)
- End Date
- Sep 10, 2026
- Fiscal Year
- 2026
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Rohrdorf (Germany)
Citations
| ID | CFR | Description |
|---|---|---|
| 23119 | ISO 13485:2016 Clause 7.3.7 | Design and development validation not performed as planned |
| 23264 | 21 CFR 820.35(a) | Lack of justification for not investigating complaints requiring investigation |
| 23267 | 21 CFR 820.35(c) | UDI was not recorded for each medical device or batch |