FDA Inspection 1336538 — Grimm Scientific Industries

Grimm Scientific Industries — FEI 1000136393

The FDA completed an inspection of Grimm Scientific Industries on Sep 18, 2026 in Marietta, OH. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 4 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
Sep 18, 2026
Fiscal Year
2026
Product Type
Devices
Project Area
Compliance: Devices
Location
Marietta, OH (United States)

Citations

IDCFRDescription
23133ISO 13485:2016 Clause 7.3.9Requirements for design and development changes not met before implementation
23138ISO 13485:2016 Clause 7.4.1Procedures for purchasing process not documented
23214ISO 13485:2016 Clause 8.2.2Complaint handling procedures not documented
23258ISO 13485:2016 Clause 8.5.2Corrective action procedures not documented

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.