FDA Inspection 538058 — The Kilroy Company

The Kilroy Company — FEI 1527214

The FDA completed an inspection of The Kilroy Company on Jan 7, 2009 in Mentor, OH. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 35 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
Jan 7, 2009
Fiscal Year
2009
Product Type
Devices
Project Area
Compliance: Devices
Location
Mentor, OH (United States)

Citations

IDCFRDescription
226921 CFR 820.20(a)Quality policy and objectives
312121 CFR 820.80(b)Lack of or inadequate receiving acceptance procedures
312521 CFR 820.80(d)Lack of or inadequate final acceptance procedures
312721 CFR 820.80(e)Documentation
315521 CFR 820.181(a)DMR device specifications
316021 CFR 820.184Lack of or inadequate DHR procedures
316421 CFR 820.184(d)Acceptance records
317221 CFR 820.198(c)Investigation of device failures
320121 CFR 820.40(a)Not approved or obsolete document retrieval
323221 CFR 820.72(a)Equipment suitability & capability
326321 CFR 820.250(b)Sampling plans
328321 CFR 820.90(a)Specific non-conforming product procedures
328521 CFR 820.90(b)(2)Product rework procedures, Lack of or inadequate procedures
328821 CFR 820.90(b)(1)Documentation of disposition, justification, signature
330021 CFR 820.100(a)(1)Statistical methodology
330421 CFR 820.100(a)(5)Changes to correct/prevent quality problems
341521 CFR 820.22Quality Audit/Reaudit - conducted
341721 CFR 820.70(a)(2)Production monitoring and control
342721 CFR 820.50(a)(2)Supplier oversight
366921 CFR 820.20(c)Management review - defined interval, sufficient frequency
368721 CFR 820.100(a)(1)Analysis of data sources
368821 CFR 820.100(a)(1)Analysis of data/reports from data sources
368921 CFR 820.100(a)(2)Investigation
369021 CFR 820.100(a)(3)Identification of actions needed
369221 CFR 820.100(a)(4)CAPA verification/validation of corrective/preventive action
369621 CFR 820.100(b)Documentation
405821 CFR 820.20Management responsibility
405921 CFR 820.22Quality Audits - defined intervals
418921 CFR 820.198(a)General
44721 CFR 820.40Lack of procedures, or not maintained
45421 CFR 820.40(a)Document review, approval by designated individual
48621 CFR 820.50(a)Evaluation of suppliers, contractors, etc., requirements
53821 CFR 820.70(a)Process control procedures
53921 CFR 820.70(b)Production and Process Change Procedures, lack of or Inad.
54621 CFR 820.75(a)Lack of or inadequate process validation