546628
Invacare Corporation — FEI 1525712
Inspection Details
- Classification
- Voluntary Action Indicated (VAI) (VAI)
- End Date
- December 19, 2008
- Fiscal Year
- 2009
- Product Type
- Devices
- Project Area
- Postmarket Assurance: Devices
- Location
- Elyria, OH (United States)
Citations
| ID | CFR | Description |
| 2371 | 21 CFR 820.30(a) | Design control - no procedures |
| 2650 | 21 CFR 820.30(f) | Design verification - Lack of or inadequate procedures |
| 3119 | 21 CFR 820.75(b) | Lack/Inad procedure-Monitoring/Control of Validated Proces |
| 3415 | 21 CFR 820.22 | Quality Audit/Reaudit - conducted |
| 3696 | 21 CFR 820.100(b) | Documentation |
| 4193 | 21 CFR 806.10(b) | Time to report - 10 days |
| 4436 | 21 CFR 803.40(a) | Report of death, injury |
| 546 | 21 CFR 820.75(a) | Lack of or inadequate process validation |
| 633 | 21 CFR 803.17(a)(3) | Lack of System for Timely Submission of Reports |
| 731 | 21 CFR 803.50(a)(1) | Report of Death or Serious Injury |
| 732 | 21 CFR 803.50(a)(2) | Individual Report of Malfunction |