FDA Inspection 568742 — Compression Therapy Concepts, Inc.

Compression Therapy Concepts, Inc. — FEI 3002834373

The FDA completed an inspection of Compression Therapy Concepts, Inc. on Mar 13, 2009 in Eatontown, NJ. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 6 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
Mar 13, 2009
Fiscal Year
2009
Product Type
Devices
Project Area
Compliance: Devices
Location
Eatontown, NJ (United States)

Citations

IDCFRDescription
237121 CFR 820.30(a)Design control - no procedures
310421 CFR 820.30(j)Design history file
315621 CFR 820.181(b)DMR production process specifications
342521 CFR 820.50(a)(1)Evaluation and Selection, Suppliers, Contractors, etc.
367421 CFR 820.30(d)Design output - documentation
367721 CFR 820.30(g)Design validation - software validation not performed

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.