FDA Inspection 602308 — Abbott Medical

Abbott Medical — FEI 2017865

The FDA completed an inspection of Abbott Medical on Jul 8, 2009 in Sylmar, CA. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 8 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
Jul 8, 2009
Fiscal Year
2009
Product Type
Devices
Project Area
Compliance: Devices
Location
Sylmar, CA (United States)

Citations

IDCFRDescription
262721 CFR 820.30(e)Design reviews performed following schedule
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
367521 CFR 820.30(f)Design verification - unresolved discrepancies
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
370421 CFR 820.250(b)Reviewed for adequacy
73121 CFR 803.50(a)(1)Report of Death or Serious Injury
73321 CFR 803.50(b)(1)Reporting Information Reasonably Known