FDA Inspection 634587 — Smiths Medical International Ltd.

Smiths Medical International Ltd. — FEI 3002088009

The FDA completed an inspection of Smiths Medical International Ltd. on Dec 10, 2009 in Watford. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 8 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Official Action Indicated (OAI)
End Date
Dec 10, 2009
Fiscal Year
2010
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Watford (United Kingdom)

Citations

IDCFRDescription
312121 CFR 820.80(b)Lack of or inadequate receiving acceptance procedures
315921 CFR 820.184DHR content
329121 CFR 820.100(b)Procedures
336921 CFR 820.198(a)(1)Uniform and timely processing
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
369621 CFR 820.100(b)Documentation
73121 CFR 803.50(a)(1)Report of Death or Serious Injury
73221 CFR 803.50(a)(2)Individual Report of Malfunction

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.