FDA Inspection 653061 — Novarad Corporation

Novarad Corporation — FEI 1000117423

The FDA completed an inspection of Novarad Corporation on Apr 2, 2010 in American Fork, UT. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 6 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
Apr 2, 2010
Fiscal Year
2010
Product Type
Devices
Project Area
Compliance: Devices
Location
American Fork, UT (United States)

Citations

IDCFRDescription
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
317221 CFR 820.198(c)Investigation of device failures
334521 CFR 820.200(a)Servicing - Lack of or inadequate procedures
367721 CFR 820.30(g)Design validation - software validation not performed
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
383721 CFR 820.25(b)Training records