FDA Inspection 659484 — Breg, Inc
The FDA completed an inspection of Breg, Inc on May 27, 2010 in Grand Prairie, TX. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 13 citations were issued during this inspection. Product area: Devices.
Inspection details
- Classification
- Voluntary Action Indicated (VAI)
- End Date
- May 27, 2010
- Fiscal Year
- 2010
- Product Type
- Devices
- Project Area
- Postmarket Assurance: Devices
- Location
- Grand Prairie, TX (United States)
Citations
| ID | CFR | Description |
|---|---|---|
| 2327 | 21 CFR 820.22 | Quality audits - Lack of or inadequate procedures |
| 3103 | 21 CFR 820.30(i) | Design changes - Lack of or Inadequate Procedures |
| 3128 | 21 CFR 820.90(a) | Nonconforming product control |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3160 | 21 CFR 820.184 | Lack of or inadequate DHR procedures |
| 3172 | 21 CFR 820.198(c) | Investigation of device failures |
| 3233 | 21 CFR 820.72(a) | Calibration, Inspection, etc. Procedures Lack of or Inadequ |
| 3669 | 21 CFR 820.20(c) | Management review - defined interval, sufficient frequency |
| 3678 | 21 CFR 820.30(g) | Design Validation - Risk analysis not performed/inadequate |
| 3837 | 21 CFR 820.25(b) | Training records |
| 4070 | 21 CFR 820.30(g) | Design validation - documentation |
| 479 | 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures |
| 630 | 21 CFR 803.17 | Lack of Written MDR Procedures |