FDA Inspection 667581 — Philips Respironics, Inc.

Philips Respironics, Inc. — FEI 2518422

The FDA completed an inspection of Philips Respironics, Inc. on Jul 1, 2010 in Murrysville, PA. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 12 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
Jul 1, 2010
Fiscal Year
2010
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Murrysville, PA (United States)

Citations

IDCFRDescription
1471021 CFR 820.150Lack of or inadequate procedures for storage.
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
230221 CFR 820.20(e)Quality System Procedures
313221 CFR 820.120Lack of or inadequate procedures for labeling
319321 CFR 820.30(g)Design validation - simulated testing
320121 CFR 820.40(a)Not approved or obsolete document retrieval
322621 CFR 820.70(g)(1)Maintenance schedule, Lack of or inadequate schedule
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
368021 CFR 820.70(a)Process control procedures, Lack of or inadequate procedures
53721 CFR 820.70(a)Production processes
63021 CFR 803.17Lack of Written MDR Procedures
73221 CFR 803.50(a)(2)Individual Report of Malfunction