FDA Inspection 667581 — Philips Respironics, Inc.
The FDA completed an inspection of Philips Respironics, Inc. on Jul 1, 2010 in Murrysville, PA. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 12 citations were issued during this inspection. Product area: Devices.
Inspection details
- Classification
- Voluntary Action Indicated (VAI)
- End Date
- Jul 1, 2010
- Fiscal Year
- 2010
- Product Type
- Devices
- Project Area
- Postmarket Assurance: Devices
- Location
- Murrysville, PA (United States)
Citations
| ID | CFR | Description |
|---|---|---|
| 14710 | 21 CFR 820.150 | Lack of or inadequate procedures for storage. |
| 14716 | 21 CFR 820.30(f) | Design verification - output does not meet input requirement |
| 2302 | 21 CFR 820.20(e) | Quality System Procedures |
| 3132 | 21 CFR 820.120 | Lack of or inadequate procedures for labeling |
| 3193 | 21 CFR 820.30(g) | Design validation - simulated testing |
| 3201 | 21 CFR 820.40(a) | Not approved or obsolete document retrieval |
| 3226 | 21 CFR 820.70(g)(1) | Maintenance schedule, Lack of or inadequate schedule |
| 3678 | 21 CFR 820.30(g) | Design Validation - Risk analysis not performed/inadequate |
| 3680 | 21 CFR 820.70(a) | Process control procedures, Lack of or inadequate procedures |
| 537 | 21 CFR 820.70(a) | Production processes |
| 630 | 21 CFR 803.17 | Lack of Written MDR Procedures |
| 732 | 21 CFR 803.50(a)(2) | Individual Report of Malfunction |