677839

Invacare Corporation — FEI 1525712

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
December 17, 2010
Fiscal Year
2011
Product Type
Devices
Project Area
Compliance: Devices
Location
Elyria, OH (United States)

Citations

IDCFRDescription
1471121 CFR 820.160(a)Lack of or inadequate procedures for distribution
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
311721 CFR 820.70(i)Software validation for automated processes
311821 CFR 820.75(a)Documentation
313021 CFR 820.100(a)Lack of or inadequate procedures
322621 CFR 820.70(g)(1)Maintenance schedule, Lack of or inadequate schedule
326621 CFR 820.86Acceptance status
328521 CFR 820.90(b)(2)Product rework procedures, Lack of or inadequate procedures
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
368021 CFR 820.70(a)Process control procedures, Lack of or inadequate procedures
369621 CFR 820.100(b)Documentation
443621 CFR 803.40(a)Report of death, injury
53721 CFR 820.70(a)Production processes
54621 CFR 820.75(a)Lack of or inadequate process validation