709082

Accurate Set Inc — FEI 2240725

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
January 28, 2011
Fiscal Year
2011
Product Type
Devices
Project Area
Compliance: Devices
Location
Newark, NJ (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
235021 CFR 820.25(b)Training - Lack of or inadequate procedures
237121 CFR 820.30(a)Design control - no procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
312121 CFR 820.80(b)Lack of or inadequate receiving acceptance procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
314921 CFR 820.180Availability
315921 CFR 820.184DHR content
316021 CFR 820.184Lack of or inadequate DHR procedures
320721 CFR 820.50(b)Supplier notification of changes
333121 CFR 820.181DMR - not or inadequately maintained
366921 CFR 820.20(c)Management review - defined interval, sufficient frequency
45421 CFR 820.40(a)Document review, approval by designated individual
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures