FDA Inspection 723738 — Pointcare Technologies, Inc.

Pointcare Technologies, Inc. — FEI 3004086182

The FDA completed an inspection of Pointcare Technologies, Inc. on Apr 28, 2011 in Marlborough, MA. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 8 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Official Action Indicated (OAI)
End Date
Apr 28, 2011
Fiscal Year
2011
Product Type
Devices
Project Area
Compliance: Devices
Location
Marlborough, MA (United States)

Citations

IDCFRDescription
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
232821 CFR 820.22Quality audits - auditor independence
312121 CFR 820.80(b)Lack of or inadequate receiving acceptance procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
319221 CFR 820.30(g)Design validation - user needs and intended uses
323321 CFR 820.72(a)Calibration, Inspection, etc. Procedures Lack of or Inadequ
326221 CFR 820.250(a)Statistical techniques - Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation