727010

North Coast Medical Inc — FEI 1000124207

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
May 25, 2011
Fiscal Year
2011
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Morgan Hill, CA (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1472221 CFR 820.40Procedures not adequately established or maintained
235021 CFR 820.25(b)Training - Lack of or inadequate procedures
316021 CFR 820.184Lack of or inadequate DHR procedures
319321 CFR 820.30(g)Design validation - simulated testing
328621 CFR 820.90(b)(1)Procedures for product review,disposition lack of/inadequate
332821 CFR 820.180(b)Retention period
341521 CFR 820.22Quality Audit/Reaudit - conducted
48621 CFR 820.50(a)Evaluation of suppliers, contractors, etc., requirements
73121 CFR 803.50(a)(1)Report of Death or Serious Injury