FDA Inspection 754176 — Innovative Health Technologies, LLC
The FDA completed an inspection of Innovative Health Technologies, LLC on Nov 22, 2011 in Salt Lake City, UT. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 16 citations were issued during this inspection. Product area: Devices.
Inspection details
- Classification
- Voluntary Action Indicated (VAI)
- End Date
- Nov 22, 2011
- Fiscal Year
- 2012
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Salt Lake City, UT (United States)
Citations
| ID | CFR | Description |
|---|---|---|
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 14722 | 21 CFR 820.40 | Procedures not adequately established or maintained |
| 2302 | 21 CFR 820.20(e) | Quality System Procedures |
| 2371 | 21 CFR 820.30(a) | Design control - no procedures |
| 3121 | 21 CFR 820.80(b) | Lack of or inadequate receiving acceptance procedures |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3160 | 21 CFR 820.184 | Lack of or inadequate DHR procedures |
| 3207 | 21 CFR 820.50(b) | Supplier notification of changes |
| 3282 | 21 CFR 820.90(a) | Nonconforming product, Lack of or inadequate procedures |
| 3285 | 21 CFR 820.90(b)(2) | Product rework procedures, Lack of or inadequate procedures |
| 3331 | 21 CFR 820.181 | DMR - not or inadequately maintained |
| 3345 | 21 CFR 820.200(a) | Servicing - Lack of or inadequate procedures |
| 3425 | 21 CFR 820.50(a)(1) | Evaluation and Selection, Suppliers, Contractors, etc. |
| 3677 | 21 CFR 820.30(g) | Design validation - software validation not performed |
| 3837 | 21 CFR 820.25(b) | Training records |
| 630 | 21 CFR 803.17 | Lack of Written MDR Procedures |