FDA Inspection 757377 — Shanghai Shunlong Physical Therapy Equipment Co. Ltd.

Shanghai Shunlong Physical Therapy Equipment Co. Ltd. — FEI 3003910843

The FDA completed an inspection of Shanghai Shunlong Physical Therapy Equipment Co. Ltd. on October 20, 2011 in Shanghai. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 7 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
October 20, 2011
Fiscal Year
2012
Product Type
Devices
Project Area
Compliance: Devices
Location
Shanghai (China)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
260421 CFR 820.30(e)Design review - Lack of or inadequate procedures
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
326221 CFR 820.250(a)Statistical techniques - Lack of or inadequate procedures
331021 CFR 820.120(b)DHR documentation of label release {see also 820.184}
331321 CFR 820.120(d)Records, DHR {see also 820.184(e)}