777228

Evoqua Water Technologies Llc — FEI 3004057358

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
April 9, 2012
Fiscal Year
2012
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Indianapolis, IN (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
235021 CFR 820.25(b)Training - Lack of or inadequate procedures
263021 CFR 820.30(e)Design review - documentation
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
313021 CFR 820.100(a)Lack of or inadequate procedures
317121 CFR 820.198(b)Rationale documented for no investigation
317221 CFR 820.198(c)Investigation of device failures
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
368021 CFR 820.70(a)Process control procedures, Lack of or inadequate procedures
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
63521 CFR 803.17(b)(1)Info evaluated to determine if event was reportable