870655
Conductive Technologies, Inc — FEI 3004539335
Inspection Details
- Classification
- Official Action Indicated (OAI) (OAI)
- End Date
- March 12, 2014
- Fiscal Year
- 2014
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- York, PA (United States)
Citations
| ID | CFR | Description |
| 14722 | 21 CFR 820.40 | Procedures not adequately established or maintained |
| 2350 | 21 CFR 820.25(b) | Training - Lack of or inadequate procedures |
| 3103 | 21 CFR 820.30(i) | Design changes - Lack of or Inadequate Procedures |
| 3104 | 21 CFR 820.30(j) | Design history file |
| 3125 | 21 CFR 820.80(d) | Lack of or inadequate final acceptance procedures |
| 3168 | 21 CFR 820.198(a) | Complaints |
| 3263 | 21 CFR 820.250(b) | Sampling plans |
| 3270 | 21 CFR 820.80(c) | Documentation |
| 3282 | 21 CFR 820.90(a) | Nonconforming product, Lack of or inadequate procedures |
| 3666 | 21 CFR 820.20(c) | Management review - Lack of or inadequate procedures |
| 3669 | 21 CFR 820.20(c) | Management review - defined interval, sufficient frequency |
| 3678 | 21 CFR 820.30(g) | Design Validation - Risk analysis not performed/inadequate |
| 3680 | 21 CFR 820.70(a) | Process control procedures, Lack of or inadequate procedures |
| 3696 | 21 CFR 820.100(b) | Documentation |
| 4059 | 21 CFR 820.22 | Quality Audits - defined intervals |
| 486 | 21 CFR 820.50(a) | Evaluation of suppliers, contractors, etc., requirements |
| 546 | 21 CFR 820.75(a) | Lack of or inadequate process validation |