FDA Inspection 877892 — Therakos, Inc.

Therakos, Inc. — FEI 2523595

The FDA completed an inspection of Therakos, Inc. on May 7, 2014 in West Chester, PA. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 8 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
May 7, 2014
Fiscal Year
2014
Product Type
Devices
Project Area
Compliance: Devices
Location
West Chester, PA (United States)

Citations

IDCFRDescription
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
312721 CFR 820.80(e)Documentation
313021 CFR 820.100(a)Lack of or inadequate procedures
315521 CFR 820.181(a)DMR device specifications
319221 CFR 820.30(g)Design validation - user needs and intended uses
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
54621 CFR 820.75(a)Lack of or inadequate process validation
73221 CFR 803.50(a)(2)Individual Report of Malfunction