FDA Inspection 885466 — Medyssey Co., Ltd.
The FDA completed an inspection of Medyssey Co., Ltd. on Jun 19, 2014 in Jecheon-si. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 7 citations were issued during this inspection. Product area: Devices.
Inspection details
- Classification
- Voluntary Action Indicated (VAI)
- End Date
- Jun 19, 2014
- Fiscal Year
- 2014
- Product Type
- Devices
- Project Area
- Postmarket Assurance: Devices
- Location
- Jecheon-si (Korea (the Republic of))
Citations
| ID | CFR | Description |
|---|---|---|
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 2604 | 21 CFR 820.30(e) | Design review - Lack of or inadequate procedures |
| 2650 | 21 CFR 820.30(f) | Design verification - Lack of or inadequate procedures |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 479 | 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures |
| 546 | 21 CFR 820.75(a) | Lack of or inadequate process validation |
| 631 | 21 CFR 803.17(a)(1) | Lack of System for Event Evaluations |