FDA Inspection 897176 — Therametric Technologies, Inc.

Therametric Technologies, Inc. — FEI 3008845669

The FDA completed an inspection of Therametric Technologies, Inc. on September 11, 2014 in Noblesville, IN. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 14 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
September 11, 2014
Fiscal Year
2014
Product Type
Devices
Project Area
Compliance: Devices
Location
Noblesville, IN (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1472221 CFR 820.40Procedures not adequately established or maintained
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
235021 CFR 820.25(b)Training - Lack of or inadequate procedures
265021 CFR 820.30(f)Design verification - Lack of or inadequate procedures
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
311921 CFR 820.75(b)Lack/Inad procedure-Monitoring/Control of Validated Proces
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
313021 CFR 820.100(a)Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
366621 CFR 820.20(c)Management review - Lack of or inadequate procedures
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
63021 CFR 803.17Lack of Written MDR Procedures