898183
AutoGenomics, Inc. — FEI 3005406097
Inspection Details
- Classification
- Voluntary Action Indicated (VAI) (VAI)
- End Date
- September 26, 2014
- Fiscal Year
- 2014
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Carlsbad, CA (United States)
Citations
| ID | CFR | Description |
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 2350 | 21 CFR 820.25(b) | Training - Lack of or inadequate procedures |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3331 | 21 CFR 820.181 | DMR - not or inadequately maintained |
| 3678 | 21 CFR 820.30(g) | Design Validation - Risk analysis not performed/inadequate |
| 3696 | 21 CFR 820.100(b) | Documentation |
| 539 | 21 CFR 820.70(b) | Production and Process Change Procedures, lack of or Inad. |