FDA Inspection 904198 — DORO INC ( O/S DORO SAFETY & SECURITY)

DORO INC ( O/S DORO SAFETY & SECURITY) — FEI 3009238346

The FDA completed an inspection of DORO INC ( O/S DORO SAFETY & SECURITY) on Nov 13, 2014 in Nepean. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 11 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Official Action Indicated (OAI)
End Date
Nov 13, 2014
Fiscal Year
2015
Product Type
Devices
Project Area
Compliance: Devices
Location
Nepean (Canada)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1472221 CFR 820.40Procedures not adequately established or maintained
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
237121 CFR 820.30(a)Design control - no procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
323321 CFR 820.72(a)Calibration, Inspection, etc. Procedures Lack of or Inadequ
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
333121 CFR 820.181DMR - not or inadequately maintained
366621 CFR 820.20(c)Management review - Lack of or inadequate procedures
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
63021 CFR 803.17Lack of Written MDR Procedures

Need the full FDA record for DORO INC ( O/S DORO SAFETY & SECURITY)? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for DORO INC ( O/S DORO SAFETY & SECURITY): recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about DORO INC ( O/S DORO SAFETY & SECURITY) →

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.