FDA Inspection 912303 — Zoll Manufacturing Corp.

Zoll Manufacturing Corp. — FEI 3008642652

The FDA completed an inspection of Zoll Manufacturing Corp. on Mar 9, 2015 in Pittsburgh, PA. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 9 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
Mar 9, 2015
Fiscal Year
2015
Product Type
Devices
Project Area
Compliance: Devices
Location
Pittsburgh, PA (United States)

Citations

IDCFRDescription
312121 CFR 820.80(b)Lack of or inadequate receiving acceptance procedures
312821 CFR 820.90(a)Nonconforming product control
316821 CFR 820.198(a)Complaints
317021 CFR 820.198(b)Review and evaluation for investigation
317221 CFR 820.198(c)Investigation of device failures
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
369621 CFR 820.100(b)Documentation
41921 CFR 820.20(b)Lack of or inadequate organizational structure
73121 CFR 803.50(a)(1)Report of Death or Serious Injury

Need the full FDA record for Zoll Manufacturing Corp.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Zoll Manufacturing Corp.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Zoll Manufacturing Corp. →

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.