FDA Inspection 923212 — MWT Materials, Inc.

MWT Materials, Inc. — FEI 3004863903

The FDA completed an inspection of MWT Materials, Inc. on April 28, 2015 in Passaic, NJ. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 6 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
April 28, 2015
Fiscal Year
2015
Product Type
Devices
Project Area
Compliance: Devices
Location
Passaic, NJ (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
310421 CFR 820.30(j)Design history file
333121 CFR 820.181DMR - not or inadequately maintained
419121 CFR 806.10(a)(1)Report of risk to health
63021 CFR 803.17Lack of Written MDR Procedures
73121 CFR 803.50(a)(1)Report of Death or Serious Injury