928601
Strata Skin Sciences, Inc. — FEI 3002942014
Inspection Details
- Classification
- Voluntary Action Indicated (VAI) (VAI)
- End Date
- June 16, 2015
- Fiscal Year
- 2015
- Product Type
- Devices
- Project Area
- Postmarket Assurance: Devices
- Location
- Carlsbad, CA (United States)
Citations
| ID | CFR | Description |
| 3103 | 21 CFR 820.30(i) | Design changes - Lack of or Inadequate Procedures |
| 3125 | 21 CFR 820.80(d) | Lack of or inadequate final acceptance procedures |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3678 | 21 CFR 820.30(g) | Design Validation - Risk analysis not performed/inadequate |
| 731 | 21 CFR 803.50(a)(1) | Report of Death or Serious Injury |
| 732 | 21 CFR 803.50(a)(2) | Individual Report of Malfunction |