932521

Transdermal Cap Inc. — FEI 3008699663

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
June 30, 2015
Fiscal Year
2015
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Highland Heights, OH (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
265021 CFR 820.30(f)Design verification - Lack of or inadequate procedures
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures