FDA Inspection 945588 — Cousin Biotech SAS

Cousin Biotech SAS — FEI 3003258310

The FDA completed an inspection of Cousin Biotech SAS on September 17, 2015 in Wervicq-Sud. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 5 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
September 17, 2015
Fiscal Year
2015
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Wervicq-Sud (France)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
260421 CFR 820.30(e)Design review - Lack of or inadequate procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation
63121 CFR 803.17(a)(1)Lack of System for Event Evaluations